Título Electromagnetic Compatibility of Medical Devices with Mobile Communications
Resumo The compatibility of mobile communication equipment with medical devices is an area where much published information and guidance is conflicting. We have hence organised a large study, based on research conducted at 18 locations including hospitals and MDA evaluation centres. Data were gathered for 178 different models of medical device using a wide range of radio handsets. Overall, in 23% of tests medical devices suffered electromagnetic interference (EMI) from handsets. 43% of these interference incidents would have had a direct impact on patient care, and were rated as serious.
Fonte Medicines and Healthcare products Regulatory Agency (MHRA)
Origem Reino Unido
Autor/Editor Medical Devices Agency (MDA)
Disponibilidade Disponível por encomenda
Endereço https://webarchive.nationalarchives.gov.uk/20141205224003/
http://www.mhra.gov.uk/Publications/Safetyguidance/
DeviceBulletins/CON007365
Referência DB 9702
Data Março 1997